Estimation of the uncertainty associated with the results based on the validation of chromatographic analysis procedures: Application to the determination of chlorides by high performance liquid chromatography and of fatty acids by high resolution gas chromatography

Producció científica: Article en revista indexadaArticleAvaluat per experts

20 Cites (Scopus)

Resum

This article presents a model to calculate the uncertainty associated with an analytical result based on the validation of the analysis procedure. This calculation model is proposed as an alternative to commonly used. bottom-up and. top-down methods. This proposal is very advantageous as the validation of the procedures and the estimation of the uncertainty of the measurement are part of the technical requirements needed in order to obtain the ISO 17025:2005 accreditation. This model has been applied to the determination of chloride by liquid chromatography in lixiviates and in the determination of palmitic acid and stearic acid by gas chromatography in magnesium stearate samples.

Idioma originalAnglès
Pàgines (de-a)107-117
Nombre de pàgines11
RevistaJournal of Chromatography A
Volum1223
DOIs
Estat de la publicacióPublicada - 3 de febr. 2012

Fingerprint

Navegar pels temes de recerca de 'Estimation of the uncertainty associated with the results based on the validation of chromatographic analysis procedures: Application to the determination of chlorides by high performance liquid chromatography and of fatty acids by high resolution gas chromatography'. Junts formen un fingerprint únic.