TY - JOUR
T1 - Validity and reliability of the satel 40 HZ stabilometric force platform for measuring quiet stance and dynamic standing balance in healthy subjects
AU - Rodríguez-Rubio, Pere Ramón
AU - Bagur-Calafat, Caritat
AU - López-De-celis, Carlos
AU - Bueno-Gracía, Elena
AU - Cabanas-Valdés, Rosa
AU - Herrera-Pedroviejo, Ernesto
AU - Girabent-Farrés, Montserrat
N1 - Publisher Copyright:
© 2020 by the authors. Licensee MDPI, Basel, Switzerland.
PY - 2020/11/1
Y1 - 2020/11/1
N2 - Background: A force platform must have validity and reliability for optimal use. The objective of this study was to analyze the validity and the reliability of the Satel 40 Hz stabilometric force platform. Methods: A study of instrumental validity and reliability, involving a cross-sectional correlational and comparative analysis was performed. To determine the validity, four certified weights located on three axes were used and the ability of the stabilometric force platform to detect changes in the position of the different axes was observed. A test–retest was performed to analyze the reliability. Forty-two symptom-free volunteers participated in the study. Assessments were taken in a standing static position and in a dynamic position, with the eyes open and closed. Three measurements were taken and the intra-class correlation coefficient (ICC) was calculated. Results: The validity increased as the weight increased for all the variables measured in the stabilometric parameters (p < 0.05). The reliability was shown to be good to excellent for the two visual conditions. The positional variables obtained a higher ICC. The variable with the best ICC was the Y mean in OE (ICC 0.874 and a p < 0.001). All the values showed an increase in a dynamic situation. Conclusion: The findings support the reliability and validity of the Satel 40 Hz stabilometric force platform. The platform could be recommended to evaluate static and dynamic standing balance in healthy adult individuals. Guidelines for treatment and the level of quality of stabilometry could be obtained from its use.
AB - Background: A force platform must have validity and reliability for optimal use. The objective of this study was to analyze the validity and the reliability of the Satel 40 Hz stabilometric force platform. Methods: A study of instrumental validity and reliability, involving a cross-sectional correlational and comparative analysis was performed. To determine the validity, four certified weights located on three axes were used and the ability of the stabilometric force platform to detect changes in the position of the different axes was observed. A test–retest was performed to analyze the reliability. Forty-two symptom-free volunteers participated in the study. Assessments were taken in a standing static position and in a dynamic position, with the eyes open and closed. Three measurements were taken and the intra-class correlation coefficient (ICC) was calculated. Results: The validity increased as the weight increased for all the variables measured in the stabilometric parameters (p < 0.05). The reliability was shown to be good to excellent for the two visual conditions. The positional variables obtained a higher ICC. The variable with the best ICC was the Y mean in OE (ICC 0.874 and a p < 0.001). All the values showed an increase in a dynamic situation. Conclusion: The findings support the reliability and validity of the Satel 40 Hz stabilometric force platform. The platform could be recommended to evaluate static and dynamic standing balance in healthy adult individuals. Guidelines for treatment and the level of quality of stabilometry could be obtained from its use.
KW - Postural control
KW - Posture
KW - Posturography
KW - Stabilometry
KW - Validity and reliability
UR - http://www.scopus.com/inward/record.url?scp=85093986136&partnerID=8YFLogxK
U2 - 10.3390/ijerph17217733
DO - 10.3390/ijerph17217733
M3 - Article
C2 - 33105835
AN - SCOPUS:85093986136
SN - 1661-7827
VL - 17
SP - 1
EP - 16
JO - International Journal of Environmental Research and Public Health
JF - International Journal of Environmental Research and Public Health
IS - 21
M1 - 7733
ER -