TY - JOUR
T1 - GCxGC-ECD
T2 - a promising method for the determination of dioxins and dioxin-like PCBs in food and feed
AU - Haglund, Peter
AU - Korytar, Peter
AU - Danielsson, Conny
AU - Diaz, Jordi
AU - Wiberg, Karin
AU - Leonards, Pim
AU - Brinkman, Udo A. T.
AU - de Boer, Jacob
PY - 2008/4
Y1 - 2008/4
N2 - There is a need for cost-efficient alternatives to gas chromatography (GC)-high-resolution mass spectrometry (HRMS) for the analysis of polychlorinated dibenzo-p-dioxins and dibenzofurans (PCDD/Fs) and dioxin-like polychlorinated biphenyls (PCBs) in food and feed. Comprehensive two-dimensional GC-micro electron capture detection (GCxGC-mu ECD) was tested and all relevant (according to the World Health Organisation, WHO) PCDD/Fs and PCBs could be separated when using a DB-XLB/LC-50 column combination. Validation tests by two laboratories showed that detectability, repeatability, reproducibility and accuracy of GCxGC-mu ECD are all statistically consistent with GC-HRMS results. A limit of detection of 0.5 pg WHO PCDD/F tetrachlorodibenzo-p-dioxin equivalency concentration per gram of fish oil was established. The reproducibility was less than 10%, which is below the recommended EU value for reference methods (less than 15%). Injections of vegetable oil extracts spiked with PCBs, polychlorinated naphthalenes and diphenyl ethers at concentrations of 200 ng/g showed no significant impact on the dioxin results, confirming in that way the robustness of the method. The use of GCxGC-mu ECD as a routine method for food and feed analysis is therefore recommended. However, the data evaluation of low dioxin concentrations is still laborious owing to the need for manual integration. This makes the overall analysis costs higher than those of GC-HRMS. Further developments of software are needed (and expected) to reduce the data evaluation time. Combination of the current method with pressurised liquid extraction with in-cell cleanup will result in further reduction of analysis costs.
AB - There is a need for cost-efficient alternatives to gas chromatography (GC)-high-resolution mass spectrometry (HRMS) for the analysis of polychlorinated dibenzo-p-dioxins and dibenzofurans (PCDD/Fs) and dioxin-like polychlorinated biphenyls (PCBs) in food and feed. Comprehensive two-dimensional GC-micro electron capture detection (GCxGC-mu ECD) was tested and all relevant (according to the World Health Organisation, WHO) PCDD/Fs and PCBs could be separated when using a DB-XLB/LC-50 column combination. Validation tests by two laboratories showed that detectability, repeatability, reproducibility and accuracy of GCxGC-mu ECD are all statistically consistent with GC-HRMS results. A limit of detection of 0.5 pg WHO PCDD/F tetrachlorodibenzo-p-dioxin equivalency concentration per gram of fish oil was established. The reproducibility was less than 10%, which is below the recommended EU value for reference methods (less than 15%). Injections of vegetable oil extracts spiked with PCBs, polychlorinated naphthalenes and diphenyl ethers at concentrations of 200 ng/g showed no significant impact on the dioxin results, confirming in that way the robustness of the method. The use of GCxGC-mu ECD as a routine method for food and feed analysis is therefore recommended. However, the data evaluation of low dioxin concentrations is still laborious owing to the need for manual integration. This makes the overall analysis costs higher than those of GC-HRMS. Further developments of software are needed (and expected) to reduce the data evaluation time. Combination of the current method with pressurised liquid extraction with in-cell cleanup will result in further reduction of analysis costs.
KW - Analysis
KW - Comprehensive two-dimensional gas chromatography
KW - Electron capture detection
KW - Food and feed
KW - Polychlorinated biphenyls
KW - Polychlorinated dibenzo-p-dioxins and dibenzofurans
UR - http://www.scopus.com/inward/record.url?scp=41149139284&partnerID=8YFLogxK
UR - https://www.webofscience.com/api/gateway?GWVersion=2&SrcApp=pure_univeritat_ramon_llull&SrcAuth=WosAPI&KeyUT=WOS:000254233900016&DestLinkType=FullRecord&DestApp=WOS
U2 - 10.1007/s00216-008-1896-0
DO - 10.1007/s00216-008-1896-0
M3 - Article
C2 - 18256808
SN - 1618-2642
VL - 390
SP - 1815
EP - 1827
JO - Analytical and Bioanalytical Chemistry
JF - Analytical and Bioanalytical Chemistry
IS - 7
ER -